TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Central axis dose calibration | At intervals not to exceed 12 months | 2 percent |
B. Constancy checks-photons | ||
(1) Dose per monitor unit along central axis | Weekly | 3 percent |
(2) Depth dose | Monthly | 2 percent |
(3) Beam uniformity | Monthly | 3 percent |
(4) Monitor chamber linearity | At intervals not to exceed 12 months | 1 percent |
(5) Timer linearity and error | At intervals not to exceed 12 months | 1 percent |
TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Field positioning aids | ||
(1) Light field and radiation field agreement | Monthly | 2 mm |
(2) Mechanical distance pins, lasers, and SSD lights | Monthly | 2 mm |
(3) Scale readouts | Monthly | 2 mm/1 degree angle |
B. Machine alignment | ||
(1) Jaw symmetry | At intervals not to exceed 12 months | 2 mm |
(2) Coincidence of collimator (jaw) and gantry axes | At intervals not to exceed 12 months | 2 mm |
(3) Stability of gantry arm and bearing under rotation | At intervals not to exceed 12 months | 2 mm |
(4) Couch motion and tabletop sag | At intervals not to exceed 12 months | 2 mm |
TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Beam uniformity | Monthly | 5 percent |
B. Depth dose | Monthly | 2 mm at therapeutic depth |
C. Dose per monitor unit constancy check | Weekly | 3 percent |
TEST TYPE | MINIMUM TEST INTERVAL | TOLERANCE |
A. Wedge transmission factor | At intervals not to exceed 12 months | 2 percent |
B. Transmission factor constancy for all treatment accessories | At intervals not to exceed 12 months | 2 percent |
*Attenuation in blocks, wedge factors, and compensator data must be checked annually. A visual inspection of the mechanical integrity of these accessories must be done monthly.
TEST TYPE | FREQUENCY | TOLERANCE |
A. Localizing lasers | Daily | 2 mm |
B. Distance indicator | Daily | 2 mm |
C. Field size indicator | Monthly | 2 mm |
D. Gantry/collimator angle indicators | Monthly | 1 degree |
E. Cross-hair centering | Monthly | 2 mm diameter |
F. Focal spot-axis indicator | Monthly | 2 mm |
G. Fluoroscopic image quality | Monthly | Established baseline |
H. Collision avoidance | Monthly | Functional |
I. Light/radiation field coincidence | Monthly | 2 mm or 1 percent |
J. Collimator rotation isocenter | At intervals not to exceed 12 months | 2 mm diameter |
K. Gantry rotation isocenter | At intervals not to exceed 12 months | 2 mm diameter |
L. Couch rotation isocenter | At intervals not to exceed 12 months | 2 mm diameter |
M. Coincidence of collimator, gantry, couch axes, and isocenter | At intervals not to exceed 12 months | 2 mm diameter |
N. Table top sag | At intervals not to exceed 12 months | 2 mm |
O. Vertical travel of couch | At intervals not to exceed 12 months | 2 mm |
P. Exposure rate | At intervals not to exceed 12 months | Established baseline |
Q. Table top exposure with fluoroscopy | At intervals not to exceed 12 months | Established baseline |
R. kVp and mAs calibration | At intervals not to exceed 12 months | Established baseline |
S. High and low contrast resolution | At intervals not to exceed 12 months | Established baseline |
MS s 144.12
32 SR 777
December 10, 2007