4731.4421 CALIBRATION OF SURVEY INSTRUMENTS.
A.
A licensee must calibrate the survey instruments used to show compliance with parts 4731.2000 to 4731.2950 and 4731.4400 to 4731.4527 before first use, annually, and following a repair that affects the calibration. A licensee must:
(1)
calibrate all scales with readings up to 1,000 millirems (10 mSv) per hour with a radiation source;
(2)
calibrate two separate readings on each scale or decade that will be used to show compliance; and
B.
A licensee may not use survey instruments if the difference between the indicated exposure rate and the calculated exposure rate is more than 20 percent.
History:
29 SR 755
Published Electronically:
March 12, 2009
Official Publication of the State of Minnesota
Revisor of Statutes