Minnesota Administrative Rules
4732.1063 WARNING DEVICES FOR INDUSTRIAL RADIOGRAPHY FACILITIES.
Subpart 1.
Open-beam configurations.
Open-beam configurations must have a readily discernible indication of:
A.
x-ray tube "on-off" status located near the radiation-producing equipment housing, if the primary beam is controlled in an "on-off" manner; or
B.
shutter "open-closed" status located near each port on the radiation-producing equipment housing, if the primary beam is controlled in "open-closed" manner.
Subp. 2.
Warning light.
An easily visible warning light labeled with the words "X-RAY ON" or other visible warning indicator that clearly shows the equipment is producing ionizing radiation, must be located near a switch that energizes an x-ray tube and illuminated only when the tube is energized.
Subp. 3.
Warning device labeling.
Warning devices must be labeled so that their purpose is easily identified.
Statutory Authority:
MS s 144.12
History:
32 SR 777
Published Electronically:
December 10, 2007
Official Publication of the State of Minnesota
Revisor of Statutes