4732.0870 REQUIREMENTS FOR STEREOTACTIC MAMMOGRAPHIC EQUIPMENT.
Subpart 1.
Equipment requirements.
Radiation-producing equipment specifically designed to perform stereotactically guided breast biopsies must meet the requirements of this chapter and:
Subp. 2.
Registrant requirements.
The registrant must ensure that:
A.
individuals operating the equipment meet the requirements of Minnesota Statutes, section 144.121, subdivision 5, or the Food and Drug Administration's Mammographic Quality Standards Act requirements;
B.
individuals have completed equipment manufacturer's training or equivalent and initial site-specific training in the registrant's operating and emergency procedures;
D.
the entire system for stereotactic breast biopsies including the equipment performance, procedures, and records are evaluated annually by a diagnostic radiographic physicist.
Subp. 3.
Quality assurance and quality control procedures.
The registrant must ensure that:
A.
all manufacturer's quality assurance or quality control procedures follow the test procedures established by the registrant, recommendations of a nationally recognized standard, or the manufacturer's specifications;
B.
the frequency of the quality assurance or quality control procedures, and corrective actions as a result of quality control testing are followed and documented; and
C.
the facility's operating and emergency procedures include quality assurance or quality control procedures.