4732.0710 RADIATION SURVEY OR MEASUREMENT INSTRUMENTS.
§
Subpart 1.
Requirements.
To ensure correct response to radiation, each radiation survey instrument must be calibrated at intervals not to exceed 24 months and after each servicing:
§
B.
be calibrated to accuracy within plus or minus 20 percent over the applicable range of the instrument;
§
D.
the calibration of any electronic equipment must be traceable to its calibration standard at the National Institute of Standards and Technology (NIST); and
§
E.
noninvasive kVp meters must be calibrated by the manufacturer or an accredited calibration laboratory.