Minnesota Administrative Rules
4731.4475 MOBILE REMOTE AFTERLOADER UNITS; ADDITIONAL REQUIREMENTS.
Subpart 1.
General requirements.
A licensee providing mobile remote afterloader service must:
A.
check survey instruments before medical use at each address of use or on each day of use, whichever is more frequent; and
Subp. 2.
Check requirements.
In addition to the periodic spot checks required under part 4731.4473, a licensee authorized to use mobile afterloaders for medical use must perform checks on each remote afterloader unit before use at each address of use. At a minimum, checks must be made to verify the operation of:
B.
source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;
G.
radiation monitors used to indicate whether the source has returned to a safe shielded position.
Subp. 3.
Simulated treatment cycle.
In addition to the requirements for checks under subpart 2, a licensee must ensure proper overall operation of the remote afterloader unit by conducting a simulated cycle of treatment before use at each address of use.
Subp. 4.
Malfunctions.
If the results of the checks required under subpart 2 indicate the malfunction of any system, a licensee must lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
History:
29 SR 755
Published Electronically:
March 12, 2009
Official Publication of the State of Minnesota
Revisor of Statutes